MicroLab PC
28 Agilent MicroLab Software Operation Manual
Generating the method file
To generate the method file in MicroLab PC:
1 From the MicroLab PC Component Definition dialog box, enter a
component name. The component name is the text that will
appear next to the value when the method is run by the end user.
The name should be descriptive of the property being predicted.
2 For a multivariate (PLS) method, choose Quant Model as the
Calculation Type. An Open file dialog box will appear.
3 Choose the appropriate calibration text file (exported from
Galactic GRAMS in Step 7 of ‘Exporting the PLS/IQ *.CAL file to
ASCII’) and click Open.
4 Other options available in the Component dialog box are listed
below:
a Peak 1 and Peak 2: These set a frequency range to be
displayed when the component is selected from the Results
screen. If this is left blank, the entire spectrum is displayed.
b Scaling: The scaling section will apply a linear scaling to the
result generated from the Quant model.
c Thresholds: These values set the display colors for the
component result. If the component value is above or below
the critical values, the result will be displayed in red. If the
component value is between the critical and the marginal
values, the result will be displayed in yellow. Otherwise the
result will be displayed in green.
d Report Value As: Changes the display between an actual
component value, a % of the critical high value or a % of the
critical low value.
5 Once all of the values are set appropriately, click OK.
6 Click Save As. Enter the appropriate method name and save to
the folder: C:\Program Files\Agilent\MicroLab PC\Methods.
7 The method is now available for use with MicroLab PC. Ensure
that the method is Activated in order to use the method.